Expand Up @@ -768,9 +768,20 @@ def initialiseLogFile():

# Check for PSS Drive Paths to save backup of Log file
if os.path.exists(R"Y:/Service/pXRF/Automatic Instrument Logs (S1Control)"):
# use gdrive path if available
# use nas path if available
driveArchiveLoc = R"Y:/Service/pXRF/Automatic Instrument Logs (S1Control)"

elif os.path.exists(R"N:/Service/pXRF/Automatic Instrument Logs (S1Control)"):
# use alt N: nas path if available
driveArchiveLoc = R"N:/Service/pXRF/Automatic Instrument Logs (S1Control)"

# this won't work, requires mounting CIFS file share and requires nas root login :(
# elif os.path.exists(
# R"smb://pss-nas.local/nas/Service/pXRF/Automatic Instrument Logs (S1Control)"
# ):
# # otherwise, check for linux smb nas access as last resort
# driveArchiveLoc = R"smb://pss-nas.local/nas/Service/pXRF/Automatic Instrument Logs (S1Control)"

if pxrf.instr_serialnumber == "UNKNOWN":
messagebox.showwarning(
"SerialNumber Error",
Expand Down Expand Up @@ -2577,17 +2588,19 @@ def getNearbyEnergies(energy, qty):


def sanityCheckSpectrum_SumMethod(
spectrum_counts: list[int], spectrum_energies: list[float], source_voltage_in_kV: int
spectrum_counts: list[int],
spectrum_energies: list[float],
source_voltage_in_kV: int,
) -> bool:
"""Checks that a spectrum is 'sensible', and that the communicated voltage is accurate.
This is required because of a 'voltage bug' in Bruker pXRF instrument software (or NSI tube firmware),
This is required because of a 'voltage bug' in Bruker pXRF instrument software (or NSI tube firmware),
sometimes causing phases of an assay to use an incorrect voltage from a previous phase.
This algorithm operates by working backwards through the list of counts (starting from the right-hand-side of the spectrum),
This algorithm operates by working backwards through the list of counts (starting from the right-hand-side of the spectrum),
and summing those counts it passes until it reaches the point where 2% of the total counts of the spectrum have been passed.
At this point on the spectrum, the energy of that channel should be below the source voltage used for the spectrum
(assuming no sum-peaks, which CAN give a false-positive). The Bremsstrahlung in the spectrum *should* drop down
At this point on the spectrum, the energy of that channel should be below the source voltage used for the spectrum
(assuming no sum-peaks, which CAN give a false-positive). The Bremsstrahlung in the spectrum *should* drop down
to near 0 counts beyond the source voltage (in keV).
`spectrum_counts` is a ordered list of 2048 integers representing the counts from each channel/bin of the detector.\n
`spectrum_energies` is a ordered list of 2048 floats representing the energy (keV) of each channel/bin of the detector.\n
`source_voltage_in_kV` is the voltage of the source for the given spectrum, AS REPORTED BY THE OEM API AND IN THE PDZ FILE.\n
Expand Down